Tools for Blot Hybridization
A mainstay of molecular biology research is the
nucleic acid blot. "Blots" are membranes such as nitrocellulose
or coated nylon to which nucleic acids have been permanently bound.
Blot hybridizations with specific nucleic acid probes provide critical
information regarding gene expression and genome structure. The
most common blot applications used in modern laboratories are Northern
blots, Southern blots and dot/slot blots. Regardless of the type
of blot, the principles of probe synthesis, hybridization, washing
and detection are the same. Ambion has invested substantial research
efforts into understanding and overcoming the limitations of each
of the steps common to all blot analyses. The following article
outlines some of these advancements and the products that have
resulted from our efforts.
Maximizing Blot Sensitivity
A primary limitation of all blot hybridizations
is the efficiency of hybridization between the nucleic acids on the
membrane, and the labeled nucleic acids in the hybridization solution.
Research at Ambion indicates that under typical blot hybridization
conditions, only 0.5-5% of the target molecules on a membrane are
actually bound by available probe. This is consistent with published
work (1). This limitation reduces the sensitivities of blot-associated
assays by 20100-fold, greatly reducing the ability to detect
rare messages with Northern blots and dot blots, or to analyze limited
DNA samples by Southern blotting. In many cases, maximizing the sensitivity
of blot hybridization would forestall the need to switch to PCR-based
detection methods, which are more difficult to set up and potentially
less quantitative.
Ambion has developed a hybridization buffer that
maximizes hybridization efficiency without increasing non-specific
background. Our ultrasensitive hybridization buffer, ULTRAhyb,
is as much as 100 times more sensitive than commonly used hybridization
buffers (Figure 1). The benefits of greater sensitivity are remarkable.
Using ULTRAhyb with Northerns, mRNAs too rare to detect using standard
means are detected in an overnight exposure. Single copy genes are
readily detected in only 1 µg of genomic DNA by Southerns only compared
to the 10 µg typically required when using standard hybridization
buffers.
|
|
|
Figure 1. ULTRAhyb Detects
a Signal in Only 8 Hours, Compared to a Minimum of 5 Days
Using Competitors' Hybridization Solutions. Six
identical Northern blots were assayed for p53 using 106 cpm/ml
of a random-primed, labeled probe. The blots were incubated
in the hybridization buffers indicated, following the manufacturer's
recommendations for time and temperature. All blots were
washed identically using 2X SSC/0.1% SDS and 0.1X SSC/0.1%
SDS. The blots were exposed to the same piece of film at
-80°C with one intensifying screen for the indicated
times.
1. Molecular Cloning: A Laboratory Manual, 1989. Sambrook,
J. Fritsch, E.F., Maniatis, T. Cold Spring Harbor Laboratory
Press.
2. Current Protocols in Molecular Biology, 1994. Ausubel,
F.M. and others, editors. John Wiley & Sons, Inc.
QuikHyb, Rapid-hyb and ExpressHyb are registered
trademarks of Stratagene, Amersham and CLONTECH, respectively.
|
Reagents for Northern and Southern Blotting
Although Ambion's positively charged nylon membranes
and wash buffers were developed primarily for Northern blot applications,
they are also well-suited for other types of blots (e.g., Southerns,
dot/slot blots, gene arrays). In addition, Ambion has recently introduced
SouthernMax Gel Prep/Gel Running Buffer and SouthernMax Pre-Transfer
and Transfer Buffers, which are specifically designed to decrease
gel running time to only 4 hrs and gel transfer time to as little
as 2 hrs. With these optimized reagents, Southern analysis can be
completed in a single day.
Transfer
NorthernMax Rapid Transfer Buffer offers
maximum transfer of RNA from agarose gels to membranes in two to
three hours, regardless of size. Ambion's NorthernMax procedure uses
downward passive transfer. Downward transfer does not crush the gel
and makes use of gravity to speed the transfer process. The alkaline
NorthernMax Rapid Transfer Buffer nicks the mRNA in the gel, reducing
even the longest transcripts to molecules that can be effectively
transferred to membrane, ensuring that all RNAs transfer with equal
efficiency.
Membranes
Ambion's positively charged nylon membrane, BrightStar-Plus,
has the lowest background and highest target retention of all membranes
that we have tested. Each lot of membrane is compared to previous
lots for consistency by our Quality Control Department, a claim that
other molecular biology companies can not make.
RNase-free Wash Buffers
Ambion's Low and High Stringency Wash buffers are
optimized to reduce background and cross-hybridization without decreasing
hybridization signal. Each lot of blot hybridization products is
functionally tested by our Quality Control Department to assure the
highest possible sensitivity and lowest background. In addition,
all solutions are certified nuclease-free.
Millennium Markers
Northern blots are used extensively to assess mRNA
size. Ambion offers Millennium Markers, a set of evenly spaced,
single-stranded RNA transcripts to help accurately determine mRNA
size. The Millennium Marker mixture includes 0.5, 1, 1.5, 2, 2.5,
3, 4, 5, 6, and 9 kilobase transcripts that can be radioactively
labeled, probed with our Millennium Marker Probe, or ethidium bromide
stained to provide an accurate sizing ladder in gels or on autoradiographs.
Compatible Reagents for Blot Hybridization
All of Ambion's blot hybridization products were
co-developed to ensure that they are completely compatible with one
another. This means that you can use our products to assemble an
optimized system that will give you the best possible results from
your Northerns, Southerns, dot/slot blots, and gene arrays.
References
- Vernier, P., Mastrippolito, R., Helin, C., Bendali, M., Mallet,
J., Tricoire, H. (1996) Radioimager quantification of oligonucleotide
hybridization with DNA immobilized on transfer membrane: application
to the identification of related sequences. Analytical Biochemistry 235,
11-19.
| Cat# |
Product Name |
Size |
| AM10100 |
BrightStar®-Plus Positively Charged Nylon Membrane (5 membranes) |
15 x 15 cm |
| AM10102 |
BrightStar®-Plus Positively Charged Nylon Membrane (small roll) |
30 x 45 cm |
| AM10104 |
BrightStar®-Plus Positively Charged Nylon Membrane (large roll) |
30 cm x 3 m |
| AM7118 |
Yeast RNA (10 mg/ml) |
10 x 1 ml |
| AM7150 |
RNA Millennium™ Markers |
25 lanes |
| AM7170 |
BrightStar® Biotinylated RNA Millennium™ Markers |
25 lanes |
| AM7785 |
Millennium™ Marker Probe Template |
10 rxns |
| AM8551 |
NorthernMax®-Gly Sample Loading Dye |
6 X 1 ml |
| AM8552 |
NorthernMax® Formaldehyde Load Dye |
6 x 1 ml |
| AM8669 |
ULTRAhyb® Ultrasensitive Hybridization Buffer |
4 x 125 ml |
| AM8670 |
ULTRAhyb® Ultrasensitive Hybridization Buffer |
125 ml |
| AM8671 |
NorthernMax® 10X Running Buffer |
1 L |
| AM8672 |
NorthernMax® One-Hour Transfer Buffer |
1 L |
| AM8673 |
NorthernMax® Low Stringency Wash Buffer |
1 L |
| AM8674 |
NorthernMax® High Stringency Wash Buffer |
1 L |
| AM8676 |
NorthernMax® 10X Denaturing Gel Buffer |
250 ml |
| AM8678 |
NorthernMax®-Gly 10X Gel Prep/Running Buffer |
1 L |
| AM9040 |
Agarose-LE™ |
100 g |
| AM9570 |
M.O.P.S. |
100 g |
| AM9680 |
Salmon Sperm DNA (sheared, 10 mg/ml) |
10 x 10 mg |
|