|
Troubleshooting
A. Low Overall Fluorescence Signal Intensity
Consider these suggestions if the average signal intensity from both your
sample RNA and the Positive Control miRNA are very low
(<500 RFU).
1. Troubleshoot miRNA
sample labeling
The most common cause of low overall fluorescence signal is poor labeling
of the miRNA sample. Troubleshooting suggestions for the mirVana miRNA Labeling Kit are provided in the Instruction Manual supplied with the product.
2. Excessive exposure of the
labeled sample to light
Fluorescent dyes are subject to photobleaching. Exposing them to light,
either before or after coupling, will reduce signal intensity. Limit the
exposure of fluorescent dyes to light by conducting coupling reactions,
labeled miRNA clean-up, array hybridization, and array washing in the
dark when possible.
3. The array slide was placed
in the scanner incorrectly
Improper placement of slides in the scanner, for example, upside down
or backwards, can result in no signal from the scanned region. Mark
your slides as you spot them so that orientation can be easily determined
later.
4. Hybridization problems
a. Incorrect ratio of miRNA Hybridization Buffer to miRNA sample
The dilution of 3X miRNA Hybridization Buffer with labeled
miRNA sample must be precise so that the final concentration is
1X miRNA Hybridization Buffer. This ensures the proper formamide
concentration for optimal hybridization stringency.
b. Hybridization and/or wash temperatures were higher than
recommended
The recommended hybridization temperature is 42°C. If in doubt,
verify the incubator temperature with a calibrated thermometer.
Conduct post-hybridization washes at room temperature. If the
ambient room temperature in your lab is over 25°C, washing may be
too stringent and could result in loss of legitimate signal.
5. Nuclease-contaminated
tubes, tips, or equipment
Using pipettte tips, tubes, or other plasticware that is contaminated
with nucleases during purification and/or handling of miRNA samples
can degrade miRNA, reducing yield and the size of miRNA. Both
RNases and DNases can be removed from surfaces using Ambion’s
RNaseZap.
B. Array Signal Not Associated with Probe
If your array image shows blotchy spots, signal that is not associated
with probe spots, and/or high background, consider the following troubleshooting
suggestions.
1. Partial or complete drying
of the hybridization
reaction on the array slide
• Exposure to air during washing is a common problem encountered
by people attempting array analysis for the first time. When miRNA
arrays are exposed to air for more than 1–2 sec, it can allow partial
drying of the array and problems with blotches, streaks, and dots of
signal that are not associated with probe.
• Make sure that the repositories in the hybridization chamber are
filled with 1X miRNA Hybridization Buffer to maintain humidity.
• Make sure that the area under the coverslip is completely filled with
miRNA hybridization mixture before starting the hybridization.
• Be sure to check the position of the coverslip over each miRNA array
just before sealing the hybridization chamber. If a coverslip gets too
close to the edge of the slide, the miRNA hybridization mixture can
wick under the slide, leaving the miRNA array with insufficient
hybridization mixture.
2. Incomplete washing
If there was a problem with washing the array, then repeating the entire
wash procedure may help remove the excess signal from the array. Following
are some precautions to observe in subsequent experiments to
avoid having to re-wash your array:
• Follow the washing procedure exactly, and be sure to completely
submerge the miRNA array in wash buffer at each step.
• Be sure to put the array slides into a clean slide rack when switching
them from the Low Stringency Wash to the High Stringency Wash.
Otherwise detergent may be carried over, causing detergent residue
on the slides.
C. High Signal from More Probes Than Expected
1. Incomplete reactive dye
quenching and removal
If the labeling moiety is not removed by column purification, it can
interact nonspecifically with elements on the array. Check the color of
the solution of the labeled and purified miRNA. The solution should be
clear or lightly tinted. More intense color likely means that uncoupled
dye is contaminating the labeled miRNA sample, and it cannot be used
for miRNA array hybridization. This could occur if too much fluorescent
dye is used in the labeling reaction, or if the glass fiber filter inside
the miRNA Labeling Column is not properly positioned and sample
bypasses the filter during the purification.
2. Low stringency
hybridization conditions
The recommended hybridization temperature is 42°C. Lower temperatures
can lead to nonspecific interactions between miRNAs in the sample
and probes on the array. If in doubt, verify the incubator
temperature with a calibrated thermometer.
3. Total RNA was used to
prepare labeled miRNA
sample
Labeling total RNA can lead to nonspecific binding to the relatively
short probes used for miRNAs. We recommend purifying miRNAs
using the flashPAGE Fractionator before labeling to eliminate
non-miRNAs from the sample before analysis with the mirVana
miRNA Array System. Alternatively miRNA can be isolated using traditional
gel electrophoresis [protocol].
|